TFM: Design of a Communication System Through the Body

The loss of a limb is a traumatic event for a human being, permanently altering the way they perform everyday tasks. Currently, most of these patients require a prosthesis for both aesthetic and functional reasons. Today, there are various types of prostheses, with robotic arms being the most commonly used for movement. These mechanical and/or electrical devices mimic human movements, controlled by the person. At present, the communication between the robotic arm and the user is established either through implants inside the body or via vulnerable communication methods.

The objective of this master’s thesis is to design a non-invasive communication system that uses the human body as a transmission medium. This approach is based on the innovative Human Body Communication (HBC) technology, which is part of Wireless Body Area Networks (WBAN). It is designed for wearable devices on the surface or inside the human body, specifically in the field of medicine. The goal is to transmit electrical or electromagnetic signals using the conductive properties of the human body.

To achieve this, an exhaustive study was conducted to understand how the human body behaves when a signal is transmitted at the selected frequency of 2.45 GHz. The signal will be transmitted through the patient’s forearm, which consists of multiple layers of materials with different dielectric properties. Therefore, computational simulations are necessary to analyze how the signal behaves.

This project is based on two experimental studies aimed at measuring propagation losses. The goal is to place the transmission and reception devices on the skin’s surface, using study [1] as a reference to calculate the signal losses through the multiple layers of the spleen. Additionally, the signal losses associated with transmission through the interior of the forearm, between the transmitter and receiver, were evaluated based on the data from study [2].

Once the system requirements were established, preliminary tests were conducted to validate the designed system. The results showed greater propagation losses than the theoretical estimates, prompting the project to focus on developing a functional prototype that is small, low-cost, and energy-efficient. The following image shows the completed system.

Finally, after testing the assembled system, it was determined that the selected insulator, RFSW-S-125-FR-PSA from Laird Technologies, is effective for communication through the air. However, aluminum yielded better results for communication through the body. On the other hand, the selected antenna was not ideal due to market limitations.

The results obtained suggest multiple areas for improvement and encourage further research to optimize the system.

References:

[1] Y. M. G. M. D. Zhi Ying Chen, «Propagation characteristics of electromagnetic wave on multiple tissue interfaces in wireless deep implant communication,» IET Microwaves, Antennas & Propagation, vol. 12, nº 13, pp. 2034-2040, 2018.

[2] C. G.-P. A. F.-L. A. V.-L. G. V. J. S. M. I. I. B. S. M. I. a. N. C. S. M. I. Ra´ul Ch´avez-Santiago, «Experimental Path Loss Models for In-Body Communications Within 2.36–2.5 GHz,» IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, vol. 19, nº 3, pp. 930-937, 2015.